Rigol DS1204B Dokumentacja Strona 77

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 132
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 76
RIGOL
Programming Guide for DS1000B Series
2-69
(6) :TRIGger:ALTernation:LEVel
Command Format:
:TRIGger:ALTernation:LEVel<value>[,<source>]
Function:
The command sets the trigger level of current channel. The <value> range
<value> range: (- 6*Scale-Offset) ~(+6*Scale+Offset).Scale is the
current vertical scale, and the unit is V/div. the <source> may be SOURceA
or SOURceB, and the source A and B is different according to the current
alternation channel.
NOTE: TtigLevel range is up to+/-6 Scale, when channel has offset, it needs to
detract offset ,such as 1V tap positon, 1V offset, the trigger range is -7V~5V.
Query Format:
:TRIGger:ALTernation:LEVel?[<source>]
Return Format:
The query returns the value of trigger voltage level, and the unit is V.
Example:
:TRIGGER:ALTERNATION:LEVEL2, SOURceB
Set the trigger voltage level as 2V.
:TRIGGER:ALTERNATION:LEVEL?SOURceB Return 2.00e+00.
(7) :TRIGger:ALTernation:EDGE:SLOPe
Command Format:
:TRIGger:ALTernation:EDGE:SLOPe<value>[,<source>]
Function:
The command sets the edge type of edge trigger in current channel as
POSitive (rising edge) or NEGative (falling edge). The <source> may be
SOURceA or SOURceB, and the source A and B are varying with the current
alternation channel.
Query Format:
:TRIGger:ALTernation:EDGE:SLOPe?[<source>]
Return Format:
The query returns POSITIVE or NEGATIVE.
Przeglądanie stron 76
1 2 ... 72 73 74 75 76 77 78 79 80 81 82 ... 131 132

Komentarze do niniejszej Instrukcji

Brak uwag