RIGOL
© 2008 RIGOL Technologies, Inc.
Programming Guide for DM3000 Series
2-16
The query returns the range parameter: 0123 or 4(see Table 2- 3).
13. :measure:voltage:AC:filter
Syntax:
:measure:voltage:AC:filter?
:measure:voltage:AC:filter{SLOW|MID|FAST}
Function:
The command queries and sets the speed of AC voltage filter.
Explanations:
:measure:voltage:AC:filter? queries the filter speed.
:measure:voltage:AC:filter{SLOW|MID|FAST} sets the filter speed as slow,
middle, or fast, and the default is fast.
Return Format:
:measure:voltage:AC:filter? returns the current filter speed: slowmid or fast.
Example:
To set the filter speed as middle:
:measure:voltage:AC:filterMID
14. :measure:voltage:AC:digit
Syntax:
:measure:voltage:AC:digit?
:measure:voltage:AC:digit{INC|DEC|5|6|7}
Function:
The command queries and sets the digit of AC voltage.
Explanations:
:measure:voltage:AC:digit? queries the digit of the AC voltage.
:measure:voltage:AC:digit{INC|DEC|5|6|7} sets the digit of the AC voltage.
The Table 2- 2 indicates the parameters and their meanings.
Remarks: If the showing digit reachs 5, DEC command will be invalid. If 7,INC
invalid.