Add 50mV for settings from >200mV/div to 10V/div
Delta Volts
Measurement
Accuracy (Average
Acquisition Mode)
Delta Volts between any two averages of 16 waveforms
acquired under same setup and ambient
conditions: ± (DC Gain Accuracy× reading + 0.05 div)
0.1div~1.0div (adjustable)
± 5 divisions from center of screen
Trigger Level
Accuracy (typical)
applicable for the
signal of rising and
falling time ≥20ns
± (0.3div × V/div)(± 4 divisions from center of
screen)
± (6% of setting + 200 mV)
Normal mode: pre-trigger (262144/ sampling rate),
delayed trigger 1s
Slow Scan mode: pre-trigger 6div, delayed trigger 6div
Set Level to 50%
(Typical)
Input signal frequency ≥50Hz
Rising, Falling, Rising + Falling
(>, <, =) Positive pulse, (>, <, =) negative pulse
Video standard &
line frequency
Support standard NTSC, PAL and SECAM broadcast systems.
Line number range: 1~525 (NTSC) and 1~625 (PAL/SECAM)
(>, <, =) Positive slope, (>, <, =) negative slope
Edge, Pulse, Video, Slope
Edge, Pulse, Video, Slope