RIGOL
© Copyright RIGOL Technologies, Inc. 2007. 1-19
User’s Guide for DS1000 Series
4. Press FORCE
Press this button starts an acquisition regardless of an adequate trigger signal,
usually used in “Normal” or ”Single” trigger mode. This button has no effect if the
acquisition is already stopped.
Key point:
Holdoff: A time interval before the oscilloscope response to next trigger
signal. During this holdoff period, the trigger system becomes “blind” to
trigger signals. This function helps to view complex signals such as an AM
waveform. Press Holdoff button to activate (
) knob, then turn it to adjust
Holdoff time.