2-22
Single Triggering
The multimeter takes one reading, or a number of readings specified by a sample
count you enter, each time press.
Press Single, the display shows:
Figure 2- 18 Set the samples
Table 2- 18 Menu Description
Function
Menu
Description
Single
Set a sample count, the default sample count is 1.
Save all changes, back to a higher level menu.
Sample Count
While the multimeter receives a single trigger signal, the multimeter takes one
reading or a number of readings.
The number of sample count range from 1 to 50,000. The factory default is 1.