Quick Guide for DS1000B Series
9
ENGLISH
5. After pressing the RUN/STOP button, the oscilloscope
does not display any waveform on screen.
Check whether the Trigger Mode is set to “Normal” or
“Single” and see whether the trigger level is out of the signal
range. If yes, set the trigger level in proper range by turning
the
knob or pressing the 50% button. Or set the
Trigger Mode asAUTO”. Moreover, push AUTO button to
display the waveform on screen.
6. After the Acquisition is set to “Averages” or Display
Persistence is set ON, the waveform refreshes slowly.
It is normal in these settings.
7. The signal is displayed as ladder like waveform.
(1) The time base setting maybe is too slow. Turn the horizontal
knob to increase horizontal resolution to improve
the display.
(2) Maybe the display Type is set to “Vectors”, and set it to “Dots”
mode to improve the display.
8. Unable to connect to PC or the PictBridge printer by
USB
(1) When connected to PC, if the message Unknown device
appears or prompting scanner or camera was found. Press
Utility to choose the I/O Setting menu, you may see the USB
Device menu has been set to be PictBridge, you should switch
it to Auto Detect or Computer. If necessary, restart the
oscilloscope.
(2) If connecting to a PictBridge printer or the printing is
unsuccessful, maybe the USB Device menu has been set to be
Computer, you should switch it to Auto Detect or
PictBridge. If necessary, restart the oscilloscope.