Rigol DS1000B Instrukcja Użytkownika Strona 158

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 157
RIGOL
User’s Guide for DS1000B Series
5-4
Trigger
Trigger Sensitivity 0.1div-1.0div (adjustable)
Trigger Level Range Internal ±6 divisions from center of screen
EXT ±1.2V
EXT/5 ±6V
Trigger Level Accuracy
(typical) applicable for
the signal of rising and
falling time ≥20ns
Internal ±(0.3div × V/div)(±4 divisions from
center of screen)
EXT ±(6% of setting + 40 mV)
EXT/5 ±(6% of setting + 200 mV)
Trigger Offset
Normal mode: pre-trigger(storage depth/(2×sampling)
rate), delayed trigger 1s
Slow Scan mode: pre-trigger 6div, delayed trigger 6div
Trigger Holdoff range 100ns~1.5s
HF Rejection 100kHz ±20%
LF Rejection 10kHz ±20%
Set Level to 50%
(typical)
Input signal frequency ≥50Hz
Edge Trigger
Edge trigger slope Rising, Falling, Rising + Falling
Pulse Trigger
Trigger condition
(, , =) Positive pulse, (, , =) negative pulse
Pulse Width range 20ns ~10s
Video Trigger
Video standard & line
frequency
Support standard NTSC, PAL and SECAM broadcast
systems. Line number range: 1~525 (NTSC) and 1~625
(PAL/SECAM)
Pattern Trigger
Pattern setup
H, L, X,
,
Alternate Trigger
Trigger on CH1, CH2,
CH3, CH4
Edge, Pulse, Video
Measurements
Cursor
Manual Voltage difference between cursors (V)
Time difference between cursors (T)
Przeglądanie stron 157

Komentarze do niniejszej Instrukcji

Brak uwag